When I increase the output amplitude on my stimulator modules during stimulation, the overload light comes on – what’s happening?

The D333H and D343 modules include an overload LED which serves to warn the user if the current requested or required cannot be supplied by the stimulator. The D333H modules have an upper limit of 1A, so if the impedance of the preparation and the voltage settings are high enough, the module will reach the maximum output current and indicate an overload. The same effect can be seen with the D343 except this is a constant current device and will supply up to 500mA until it reaches a compliance limit of 100V. If you run into the above problems, consider whether you need to have the output settings so high and also check that your preparation and electrodes provide a low impedance connection.