In external voltage control mode I am seeing values on the current display that fluctuate significantly even if my input voltage is at zero. How can I prevent this?
When using the analogue voltage control, the DS8R uses the 0-10V input to set the output current range of 0-1000mA and as a result, a fluctuation of just ±1mV in the input voltage signal will influence the current output by ±100µA. If you short circuit the analogue input BNC socket of the DS8R you can see if the fluctuation you are seeing in the output is due to noise within the DS8R itself or from the NI-DAQ or BNC cable.
The best way to reduce the effect of this noise in the external voltage signal, would be to set the DS8R range to an alternative value more appropriate for your needs. For instance, if you only stimulate over the range of 0-10mA, set the DS8R to 10mA, press and hold the amplitude dial and then switch to the external voltage control mode. The DS8R display will show that a 0-10V input has been re-scaled and will now correspond to 0-10mA at the output. Re-zeroing at this point should result in a steady 0mA on the display confirming that the noise in the input signal is no longer having a significant effect on the output current.