I would like to stimulate neurons using 20V 1ms long pulse applied at 10Hz frequency for 30 seconds. Such train should be programmed elsewhere as a train of TTL pulses and delivered to the stimulator, am I right?
Correct, you should program your DAQ interface to deliver trigger pulses @ 10Hz for 30s via a digital output. The DS2A will trigger on each rising edge and the stimulus duration will be set to 1ms using the controls on the DS2A front panel, with the voltage set to 20V.