I understand the TVEC catheters are supplied for single use once sterilised. Can you tell me why they cannot be re-sterilised and re-used?
Due to the coiled construction of the electrodes, it is unsafe to attempt to reprocess the device for reuse. During use the coil is flooded with blood including any blood-borne pathogens and proteins. Subsequent cleaning cannot provide the mechanical actions required to necessarily remove coagulated or dried blood from behind the coil. Further exposure to sterilisation processes cannot be guaranteed to achieve the kill level required to attain a Sterility Assurance Level of 10-6 thereby exposing the animal to risk of infections.