I understand the D360R is based on the 8 Channel D360, other than the number of channels, how do they differ?
The D360 is certified to the European Medical Device Directive and is suitable for patient use, while the D360R is designed for safe use in human research, but is not a medical device. The D360 incorporates electrode impedance checking via the control software and headstage LEDs; the D360R does not have any electrode impedance checking mode. The D360 allows for internal linking of +ve or -ve electrode inputs, so multiple channels can share a single reference electrode. The D360R operates in standard differential mode with an additional Common input. The D360 software allows for multiple groups of channels. The D360R allows channels to be linked and share settings, but does not permit multiple channel grouping.