I recently had a D185 MultiPulse Cortical Stimulator serviced and as part of this service the output transformer was replaced. I used to be able to set this D185 to 600V and 5 pulses, but now the unit trips at these settings and displays the ERROR warning. What is wrong?

The symptoms most likely reflect the accepted tolerances in the voltage/frequency (Vf) trip of individual D185’s. As this particular unit had the transformer replaced during its last service, this characteristic of the unit would have been altered and this is not something that can be calibrated during a service.
If you examine the graph on p37 of the operators manual where the red lines are from an example D185 unit it is possible to see that with a voltage set to 600V and with a pulse train of 5, the output voltage could lie close to the black line indicating when the safety trip would occur. Our test specification allows for a 5% tolerance in this output, which could lead to an actual measured voltage of between 570V and 630V. It is most likely that the replacement transformer gives a slightly higher voltage when set to 600V and so triggers the safety Vf trip of the D185 earlier than before it had a service.