I have recently started using a mixture of recording electrode types and now I am having difficulty seeing the resultant EEG activity. It appears as if the amplifier is blocked, as the de-block function momentarily brings the signal back. What is happening?

It is possible that the problem is due to the difference in materials that the electrodes are constructed from. The D360 circuitry is designed for electrodes which are at similar potentials (i.e an input offset different by no more than 100mV). If you use a mixture of electrodes types it is possible that the input offset would exceed 100mV and the amplifier would not be able to compensate for this difference in potential. We would recommend that you determine the potentials the electrodes are at to see if that could be an explanation and consider reverting back to electrodes constructed from the same materials.