I have enabled the “D360 signal conditioner” feature in the CED software, but should I also run the D360 Client Software?

The latest versions of CED Spike2 and Signal software packages now have the ability to control the gain and filter settings of the D360 amplifier as well as check the D360 channel impedance values. This means that gain and filter settings are stored alongside the recorded data and changes can be made without having to leave the acquisition software. However, it should be noted that if you run the D360 Client software simultaneously, both it and the CED software will have access to make changes to amplifier settings and this could become confusing. We therefore suggest that Signal and Spike2 users do not run the D360 Client Software and make all changes to D360 settings within the CED software.