I am intending to use the D185 Mark II with a D185-HB3 electrode connection headbox for an upcoming scoliosis surgery but I was wondering how the polarity switches on the stimulator and the one on the D185-CB1 will interact with each other?
When the D185 Mark II and the D185-CB1 are connected, both methods of reversing the stimulus polarity will operate. For example if both the D185 and D185-CB1 are set to reversed polarity, then the output will have NORMAL polarity. Digitimer suggest that if you are using the D185 Mark II along with a D185-CB1, you should leave the D185 front panel switch set to normal polarity (flick switch to the left) and only use the D185-CB1 to change it.