I am considering using the DS5 for some studies of rodent nerve excitability. Is the DS5 suitable for this?

The DS5 is designed for use with human nerves in vivo and as such the sort of currents it generates are in the mA range. Even when the output of the DS5 has been recently autozeroed, some small currents (in the µA range) can persist and while these would be insignificant for human studies, they may be unwelcome in the rodent equivalent. Instead, we would recommend our DS4 stimulator for such non-human research studies.