How does the DS7R differ from the DS7A and DS7AH?

The DS7R has the same maximum current output as the DS7AH, but the more flexible pulse duration settings of the DS7A (50µs to 2ms), making it more versatile than the DS7AH. However, because the maximum energy that the stimulator can deliver is in excess of 50mJ, which exceeds the EN60601-2-40 standard for evoked potential stimulators, we are not able to give the DS7R a medical CE mark against that standard. Instead, the DS7R is offered for human research applications, where more flexible pulse parameter settings may be required.