As I want to continue to record SEPs while stimulating transcranially with the D185, I am worried that the stimulus artefact from the D185 will saturate my EP system inputs? Is there any way to prevent this from happening?

Yes, the D185-HB3 was specifically designed with this problem in mind. The D185-HB3 headbox incorporates 5 pairs of MEP output sockets as well as 6 channels of isolation for SEP electrodes. If the SEP electrodes are connected through the D185-HB3, they are briefly isolated while an MEP stimulus is passed. This prevents the EP system from being exposed to the MEP stimulus artefact and thus stops the amplifier from saturating/blocking for any length of time. You can read more about the D185-HB3 on the D185 page.