1.  I am using the D360 Amplifier but I do not have a users manual for it - Is it possible to obtain a copy from Digitimer?

2.  I want to obtain the latest version of the D360 control software, where can I download it?
3.  My electrode leads terminate with 2mm plugs which are not compatible with the touch-proof sockets on headbox of the D360, is it possible to purchase appropriate adaptors?
4.  I have recently started using a mixture of recording electrode types and now I am having difficulty seeing the resultant EEG activity.  It appears as if the amplifier is blocked, as the de-block function momentarily brings the signal back.  What is happening?
5.  I want to upgrade to the most recent D360 Control Software, but I am not sure it is compatible with my amplifier, how do I tell?
6.  I want to use my D360 with a laptop computer, but the computer does not have a serial port, what can I do?
7.  I have installed a USB/Serial adaptor supplied by Digitimer, but my computer is failing to communicate with the D360.  Can you help?

 

1.  I am using the D360 Amplifier but I do not have a users manual for it - Is it possible to obtain a copy from Digitimer?

Yes, the users manual is available as a PDF embedded in the D360 Control Software.  Just click on "Help" within the menu bar and you can open a copy and save it for future reference..

 

2.  I want to obtain the latest version of the D360 control software, where can I download it?

The latest version of the software can be downloaded from the D360 Software page of this website.

 

3.  My electrode leads terminate with 2mm plugs which are not compatible with the 1.5mm DIN touch-proof sockets on headbox of the D360, is it possible to purchase appropriate electrode adaptors?

Yes, Digitimer has these adaptors in stock (see photo below), please contact us for current information on availability and pricing of part code E.A0110.

  

  

4.  I have recently started using a mixture of recording electrode types and now I am having difficulty seeing the resultant EEG activity.  It appears as if the amplifier is blocked, as the de-block function momentarily brings the signal back.  What is happening?

It is possible that the problem is due to the difference in materials that the electrodes are constructed from.  The D360 circuitry is designed for electrodes which are at similar potentials (i.e an input offset different by no more than 100mV).  If you use a mixture of electrodes types it is possible that the input offset would exceed 100mV and the amplifier would not be able to compensate for this difference in potential.  We would recommend that you determine the potentials the electrodes are at to see if that could be an explanation and consider reverting back to electrodes constructed from the same materials.

  

5.  I want to upgrade to the most recent D360 Control Software, but I am not sure it is compatible with my amplifier, how do I tell?

The most up to date version of the software is available here, however, it must not be used with D360 amplifiers of serial numbers lower than 51, unless they have been specifically modified by Digitimer to accept this version.  Please note that the current version of the D360 software is NOT Windows 95 or 98 compatible.

  

6.  I want to use my D360 with a laptop computer, but the computer does not have a serial port, what can I do?

Digitimer have successfully tested and can supply a type of USB to serial port adaptor which allows the D360 to be run via a USB port.  However, not all USB to serial port adaptors are compatible, so please contact us further information.

7.  I have installed a USB/Serial adaptor supplied by Digitimer, but my computer is failing to communicate with the D360.  Can you help?

Typically this problem is due to the USB/serial adaptor not being recognized by Windows or the assigned COM port being to high.  Please check within the Device Manager that it is installed correctly and if not, reinstall it using the VCP driver software from the FTDI website.  It should also be noted that the Digitimer D360 Control software needed to be installed using an administrator account.

If the installation of the software and adaptor has completed successfully, but the D360 is still not recognised by the computer, check that the COM port assigned to the USB Serial adaptor is lower than 10.  COM port values higher than 10 are not visible to the D360.  Changes to the COM port assignment can be made within the "USB Serial Adaptor" section of the Windows Device Manager.


HOME > SUPPORT > CLINICAL FAQ's > D360


[Latest News]    [Newsletter]    [Downloads]    [Contact Us]    [Support]    [Search]   [Website Guide]    [Privacy]    [Company Information]

 

© 2011 Digitimer Limited - "Digitimer" and "NeuroLog" are registered Trade Marks of Digitimer Ltd.
Product Specifications are subject to change without prior notice
Subscribe to E-News                   Read about how we use cookies

Join us on Facebook